The EF is a class II biological safety cabinet. The EF offers the user the possibility to work safely with bioagents under sterile conditions, via the work opening, without complete physical shielding of the experiment.
Average downflow (in working position): 0.36 m/s ±0.05
Average downflow (in standby position): 0.10 – 0.20 m/s
Average inflow at the top of the working opening (in working position): 0.28 m/s
Average inflow through the entire working opening (in working position): > 0.5 m/s
Air exhaust (in working position): 400 m³/h ± 25
Sound level: < 57 dB A
Lighting: 2x TL-high output (white – No. 840)
Lighting intensity: 1000 Lux
UV lighting type: UV-C
UV lighting wavelength: 253 nm
Pre-filter: G3 (EN 779) (1150 x 445 mm)
Downflow filter: Astrocell HEPA 12P6+ (H14 acc. EN 1822) (610 x 1220 x 69 mm)
Emission filter: Astrocell HEPA 4P4 (H14 acc. EN 1822) (457 x 457 x 69 mm)
Working opening in working position (HxW): 165 x 1190 mm
Maximum working opening: 350 mm
Working surface (WxD): 1109 x 440 mm
Dimensions (WxDxH): 1333 x 844 x 1552 mm
Height including base: 2297 mm
Weight: 255 kg
Weight | 0,0 kg |
---|---|
Warranty | 6 months |
Condition | Used, in good condition |
Date of Manufacture | 2016 |
Brand | Telstar |